Accreditation Scope
Search, filter and download which instruments are calibrated, over which ranges and with which uncertainties (CMC).
The current official scope is published on the TÜRKAK website. This list is for information; last updated: 28.09.2026
20 records listed
| Instrument | Method | Measurement Range | CMC (k=2) | Location |
|---|---|---|---|---|
| Dimensional Calibration | ||||
| Digital / Analog Caliper A | DIN 862 / KM-B-01 | (0 … 1000) mm | (10 + 12·L) µm, L: m | Lab + On-site |
| Outside Micrometer A | ISO 3611 / KM-B-02 | (0 … 500) mm | (1,5 + 5·L) µm | Laboratory |
| Dial Indicator A | ISO 463 / KM-B-03 | (0 … 100) mm | 2,2 µm | Laboratory |
| Gauge Block A | ISO 3650 / KM-B-04 | (0,5 … 100) mm | (0,06 + 0,8·L) µm | Laboratory |
| Profile Projector A | KM-B-07 | (0 … 200) mm | 3 µm | On-site |
| Temperature Calibration | ||||
| Digital Thermometer (Probe) A | EURAMET cg-8 / KM-S-01 | (-80 … 660) °C | 0,05 °C | Laboratory |
| Thermocouple (K, J, N, S) A | EURAMET cg-8 / KM-S-02 | (0 … 1200) °C | 1,2 °C | Laboratory |
| Oven / Furnace Mapping A | DKD-R 5-7 / KM-S-05 | (-40 … 1000) °C | 0,6 °C | On-site |
| Pressure Calibration | ||||
| Pressure Gauge (Hydraulic) A | EURAMET cg-17 / KM-P-01 | (0 … 1000) bar | 0,015 % okuma | Lab + On-site |
| Pressure Transmitter A | EURAMET cg-17 / KM-P-02 | (-0,95 … 700) bar | 0,02 % FS | Laboratory |
| Vacuum Gauge A | KM-P-04 | (-0,95 … 0) bar | 0,5 mbar | Laboratory |
| Mass & Balances | ||||
| Electronic Balance (Class I–II) A | EURAMET cg-18 / KM-K-01 | (0 … 60) kg | 0,02 mg … 0,4 g | On-site |
| Reference Weight (F1–M1) A | OIML R111 / KM-K-02 | (1 mg … 20 kg) | 0,003 mg … 25 mg | Laboratory |
| Electrical Calibration | ||||
| Digital Multimeter A | EURAMET cg-15 / KM-E-01 | DC: (0 … 1000) V | 12 µV/V + 2 µV | Laboratory |
| Clamp Meter A | KM-E-03 | AC: (0 … 1000) A | 0,35 % | Laboratory |
| Insulation Tester A | KM-E-05 | (1 MΩ … 10 GΩ) | 0,5 % | Laboratory |
| Force & Torque | ||||
| Torque Wrench A | ISO 6789-2 / KM-T-01 | (1 … 1500) N·m | 0,5 % | Laboratory |
| Force Gauge A | ISO 376 / KM-T-03 | (10 N … 50 kN) | 0,1 % | Laboratory |
| Hardness Calibration | ||||
| Rockwell Hardness Tester A | ISO 6508-2 / KM-H-01 | (20 … 70) HRC | 0,6 HRC | On-site |
| Portable Leeb Hardness Tester A | ISO 16859-2 / KM-H-04 | (300 … 900) HLD | 6 HLD | Lab + On-site |
A Service within accreditation scope. CMC: calibration and measurement capability, expanded uncertainty at ~95% confidence (k=2).